Lab Testing

HPLC, Mass Spec, Endotoxin: What the HHS Peptide Testing Standard Asks a COA to Show

The HHS peptide policy includes a testing standard for bulk peptide APIs used in compounding. What each test measures (HPLC assay, mass spec identity, peptide content, endotoxin, residual solvents, TFA), what the COA must state, and how it compares with typical research-market reports.

Key takeaways

  • The HHS peptide policy (30 October 2026) sets out a “Testing Standard for Peptides Used in Compounding”. Condition 3 requires the bulk API’s COA to include results consistent with it.
  • The standard separates identity (at least two independent, “orthogonal” methods, such as mass spec plus amino-acid analysis), assay (HPLC against a reference standard), peptide content, impurities and contaminants (endotoxin, bioburden, solvents, metals, TFA, water).
  • The COA must also name the original manufacturer and its FDA Establishment Identifier (FEI), the batch, the synthesis and release dates, the grade “Intended for human drug compounding”, a CGMP statement and a quality-unit signature.
  • Many research-market COAs report HPLC purity, MS identity and net content, and sometimes endotoxin. That covers part of the standard, not the whole of it. A “99% purity” figure alone does not answer most of these questions.

What testing does the HHS peptide policy require?

Condition 3 of the HHS policy requires the peptide bulk drug substance to come from an FDA-registered facility and to be “characterized and identified in a valid certificate of analysis (COA) that includes test results consistent with existing guidance-based criteria described by FDA”. HHS says the point is assurance that what a compounder receives “is what it claims to be and has been tested for identity, purity, potency, and the absence of harmful contaminants such as endotoxins, heavy metals, microbial contamination, and impurities” (HHS policy).

The policy attaches a testing table that lists each quality attribute, acceptable analytical methods, acceptance criteria and USP or ICH references. A second table lists acceptable “orthogonal” methods. This article explains what those tests measure in plain terms. It applies to bulk API for licensed 503A compounders, and it is also a useful reference for reading any peptide COA.

What does “orthogonal methods” mean?

The standard defines orthogonal methods as “at least two independent analytical procedures relying on distinct physicochemical principles” (HHS, footnote 3). It requires them for physicochemical properties, identity and impurities, and for larger peptides (eight amino acids or more) for higher-order structure, aggregates and biological activity.

Different methods have different blind spots. Mass spectrometry confirms that a molecule of the right mass is present, but it cannot always tell two sequences with the same composition apart. Amino-acid analysis confirms composition but not order. Tandem MS sequencing, Edman degradation or peptide mapping confirm order. Two methods that work on different principles make a misidentification much less likely.

Which tests are in the standard, and what does each one measure?

AttributeMethods namedPlain-English meaning
Identification / primary sequenceTwo orthogonal methods from: HPLC co-elution with a reference standard, MS (monoisotopic mass), MS/MS sequencing, quantitative amino-acid analysis, NMR, enantiomeric purity, Edman, peptide mapping. “100% sequence coverage”Is this the right molecule, with the right amino acids in the right order and the right chirality?
Assay (drug substance)HPLC / RP-HPLC against an established quantitative reference standard (USP <621>)How much of the labelled peptide is there, measured against a known standard rather than as a percentage of peaks
Peptide content (net)Amino-acid analysis, nitrogen analysis, UV spectroscopy or quantitative NMRHow much of the powder’s weight is peptide, as opposed to salt, water and counter-ions
Peptide-related impuritiesOrthogonal methods such as LC-MS/MS, UHPLC-HRMS, HILIC-MS, CE-MS; limits per USP <1503> / ICH Q3ATruncated, deleted, oxidised or otherwise altered versions of the peptide
Nitrosamine impuritiesPer FDA nitrosamine guidance, when applicableA class of potentially carcinogenic impurities FDA controls in all drugs
Residual solventsGas chromatography (USP <467>, ICH Q3C)Leftover synthesis solvents
Elemental impuritiesICP-MS (USP <232>/<233>, ICH Q3D)Heavy metals and catalyst residues
Residual TFA / fluorideIon chromatography / ion-selective electrodeTrifluoroacetic acid left from synthesis and purification
Counter-ion (acetate) contentUSP <503>Confirms the declared salt form, which the policy specifies for each peptide
Water contentKarl Fischer titration (USP <921>)Moisture in the lyophilised powder
BioburdenMicrobial enumeration and specified organisms (USP <61>, <62>)Live microbial contamination
Bacterial endotoxinsBacterial endotoxins test, e.g. LAL (USP <85>)Fever-causing fragments of bacteria, which sterilisation does not remove
Larger peptides (≥8 amino acids)Higher-order structure (CD, NMR, FTIR), aggregates (SEC, SEC-MALS, AUC, DLS), bioassays, in-vitro immunogenicity assays (ELISA, SPR)Folding, clumping, biological activity and immune-reaction risk

Source for every row: the HHS testing standard and its Table 1 (HHS policy). HHS notes that bioburden and endotoxin are critical quality attributes for injectables and that other dosage forms need tests appropriate to that form.

Why is “99% purity” not the same as “99% peptide”?

This is the most common misreading of a peptide COA. An HPLC purity figure is usually the share of the chromatogram’s peak area that belongs to the main peak. It shows how much of what the detector saw was the target peptide, compared with peptide-related impurities. It does not show:

  • Net peptide content. A lyophilised peptide powder also contains counter-ions (acetate or TFA) and water. That is why the standard lists peptide content and counter-ion, TFA and water testing as separate attributes.
  • Assay against a reference standard. The standard asks for HPLC compared with an established quantitative reference standard, which gives an amount, not just a ratio.
  • Anything the HPLC detector cannot see. Endotoxins, microbes, metals and residual solvents need their own tests.

A vial can therefore show “99% purity” and still differ from its label quantity, carry a different salt form, or fail an endotoxin limit. The COA has to report those things separately for a reader to know.

What must the COA itself state?

The standard’s footnote 1 lists the information required on the API documentation (HHS policy):

  • Bulk drug substance name (USAN, INN or IUPAC), structure, amino-acid sequence, molecular formula and weight
  • Manufacturer’s corporate name, FDA Establishment Identifier (FEI) and address, plus any processing firm and repacker, and the testing laboratory
  • Batch number, date of synthesis, date of release, expiry or retest date, and storage conditions
  • Grade: “Intended for human drug compounding”, and the intended dosage form
  • A statement of CGMP compliance, dated and signed by the quality unit
  • For each test: the method name and number, and the method validation or verification report number, plus the reference standards used

Condition 4 adds that the physical location of the original manufacturer’s facility must be on the COA. A repacker or reseller must have validated both the manufacturer and the manufacturer’s COA. McDermott notes that this reaches into the supply chain beyond section 503A’s existing registration requirement (McDermott).

How does this compare with a typical research-market COA?

ElementOften on research-market COAsHHS compounding standard
HPLC purityCommonRequired as assay against a reference standard, plus impurity profiling
Mass-spec identityCommonOne of at least two orthogonal identity methods, with full sequence coverage
Net content / quantitySometimesPeptide content by AAA, nitrogen, UV or qNMR
Endotoxin, microbialSometimesRequired for injectables (USP <85>, <61>/<62>)
Heavy metalsOccasionallyICP-MS per USP <232>/<233>
Residual solvents, TFA, water, counter-ionRareEach required
Manufacturer FEI, address, CGMP statement, quality signatureRarely; RUO reports usually name the client, not the manufacturerRequired

The middle column reflects the report formats we see when indexing labs in our testing facilities directory. It is a qualitative description, not a count. A research-market report can be accurate and still answer a narrower question than a compounding COA. Step-by-step checks for any COA are in how to verify a peptide COA, and what to publish on a COA page is in COA pages.

Who is this standard for?

Formally, it applies to the bulk peptide API that a 503A pharmacy, federal facility or physician uses under the HHS policy. It is the compounder’s condition to meet, and in practice the API manufacturer’s documentation to supply. It does not regulate research-use-only sellers, who are outside the policy altogether (RUO vs compounded).

It is also the first detailed public statement from HHS of what a fully characterised peptide API looks like. Before this, FDA’s category 2 notes for these peptides repeatedly cited “complexities with regard to peptide-related impurities and API characterization” (FDA). Testing labs, suppliers and buyers now have a public reference to compare reports against.

Sources

Checked 4 November 2026. This is an explainer on analytical testing. It is not a quality certification of any product and not medical advice.

Reader Question

“Our COA shows HPLC and mass spec. Is that enough under the new standard?”

Short Answer

For a compounding API under the HHS policy, no. The standard adds a second identity method, assay against a reference standard, peptide content, impurities, solvents, metals, TFA, water, counter-ion and microbial tests, plus manufacturer details. For research-market reports, HPLC and MS answer the narrower purity and identity questions.

Practical Recommendation

  1. Read purity, identity and content as three separate results. Do not let one “99%” figure stand in for all of them.
  2. For any compounding API, check that the COA names the original manufacturer, its FEI number and address, and carries a quality-unit signature.
  3. For injectable forms, look for endotoxin (USP <85>) and bioburden results on the same lot.

What we learned

When we index testing labs and supplier COAs, the gap usually isn’t a false number. More often the report answers a narrower question than the reader assumes. A clean HPLC trace and a matching mass are real results. Salt form, net content, endotoxin and the manufacturer’s identity are separate questions, and the HHS standard is the first public checklist that lists them all side by side.

Frequently asked

What is an orthogonal test in peptide analysis?

Two or more independent methods based on different physical or chemical principles, such as mass spectrometry and amino-acid analysis, used together so one method’s blind spots are covered by the other.

Does 99% HPLC purity mean the vial holds 99% peptide?

No. HPLC purity compares the main peak with other detected peaks. Water, counter-ions such as acetate or TFA, and contaminants like endotoxin need separate tests, and peptide content is measured separately.

What is an FEI number?

FDA’s Establishment Identifier for a registered facility. The HHS standard asks the COA to give the manufacturer’s FEI and address.

Why does endotoxin testing matter for peptides?

Endotoxins are bacterial fragments that can cause fever and other reactions and are not removed by sterilisation. The HHS standard treats endotoxin testing per USP <85> as a critical attribute for injectable forms.

Do research-use-only suppliers have to meet this standard?

No. It applies to bulk API used by licensed 503A compounders under the HHS policy. It is still a useful benchmark for reading any peptide COA.